Low ROX ABI: 7500, 7500 Fast, ViiA 7™, QuantStudio™
Quantstudio 3 requires low ROX
-
-
www.thermofisher.com www.thermofisher.com
-
while not required, using a passive reference like ROX dye to normalize data helps achieve a higher level of precision among technical replicates. Without normalization, more replicates may be required to achieve comparable precisions levels, therby increasing the time and resources
ROX increases precision of technical replicates (which we don't do much of in our lab)
-
ROX fluorescence is affected by anything else that would alter overall fluorescence readings, such as: • Bubbles in wells • Evaporation • Condensation or droplets • Instrument issues, such as electrical surges
ROX helps normalize these variations
-
-
signagen.com signagen.com
-
ROX is normalising for the attributes of the qPCR mixture, such as volume differences, presence of bubble etc.
passive reference ROX helps with these
-
differences in reaction volumes due to pipetting errors can bias the results. Also, issues associated with the qPCR instrument, such as the optics, can also influence the fluorescent signal generated.
passive reference ROX helps with these
-